Add to Cart
Features:
Ø 3.5 inch LCD screen, multiple languages customization
Ø Wide range of flow rate:0.01-2000ml/h
Ø Comprehensive audible and visual alarm function, adjustable volume, visible from 360°
Ø Ultrasonic bubble detection technology
Ø Change flow rate without stopping the current task
Ø Optional: thermostat function, drop sensor
Model No. Parameters | DY-113 | |||
Infusion Principle | Fingertip Peristaltic Pump | |||
I.V.Set Compatibility | All brands | |||
Range of Flow Rate | 0.01~1800ml/h (1000-1800ml / h by A.C.) | |||
Drug Name Display | Default 12 drugs, can be updated & added to 2000 | |||
Infusion Rate Step By | Increments at 0.01 ml/h | |||
Infusion Mode | Time Mode, Rate Mode, Body Weight Mode mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc. | |||
Purge/Bolus Rate | 1-1800ml/h, can be programmed by user,default value 800ml/h (1000-1800ml / h by A.C.) | |||
Time Preset | 0 h : 0 min~9999h : 59min | |||
Volume Range | 0.01~9999.99ml | |||
Accuracy | ±5% | |||
Volume Infused | 0-9999.99ml | |||
Total Volume Infused | 0-999999ml | |||
Bubble Sensitivity | Detestability of air bubble ≥40uL | |||
Occlusion Pressure | High: 100KPa±30KPa | |||
Middle: 60 KPa±25KPa | ||||
Low: 40 KPa±20KPa | ||||
Visible & Audible Alarms | Air in line, Occlusion, Door open, VTBI completion, Low battery,,Drop Error,Pause Overtime, Master CPU Malfunction,Monitoring CPU Malfunction,Air Detect Module Error,Pressure Detect Module Error,Motor Malfunction,Battery Exhausted, , Heating Error , A.C Connection, A.C Disconnection | |||
K.V.O rate | 1ml/h~5ml/h, defaul value 1ml/h, can be programmed by user | |||
Internal Battery | DC:10.8V,Lithium battery support from 7 to10 hours | |||
Power | AC:100V-240V 50HZ/60HZ | |||
Power Consumption | 30VA | |||
Working Environment | Temperature: +5--+40°C ;Atmospheric pressure: 20%--90%; Relative Humidity: 700hPa--1060hPa | |||
Classification | Class II,Type CF, IPX4 | |||
Dimension & Weight | 140mm(width)×260mm(length)×110mm(height) ; 2.5Kg | |||
Optioanl Function | Event Log with at least 2000 history records,DC, Heating Function |